The LPMIE Pilot Vacuum Freeze Dryer with Stoppering System is designed for pilot-scale freeze-drying of vial products. It supports the processing of bulk materials and vial samples, with a dedicated stoppering mechanism for sealing vials after the drying process.
Before freeze-drying, samples are filled into vials and the stoppers are partially inserted. After the freeze-drying cycle, the stoppering system presses the stoppers into place inside the chamber. This helps reduce exposure to the surrounding environment and moisture reabsorption during product handling.
The equipment is suitable for pharmaceutical, biotechnology, and research applications involving vial-based products.
l In-Chamber Vial Stoppering: Vials can be sealed inside the chamber after freeze-drying.
l Vial and Bulk Material Processing: Supports vial samples as well as liquid, paste, and solid materials, subject to process configuration.
l Silicone Oil Shelf Heating: Provides controlled shelf temperature during the freeze-drying cycle.
l Optional Process Monitoring: Options include eutectic point testing, endpoint detection, data recording, and vacuum monitoring.
l Flexible Configuration: Additional functions can be selected according to product and process requirements.

l Pharmaceutical vial products
l Injectable formulations and freeze-dried powders
l Biological products and diagnostic reagents
l Biotechnology research samples
l Pilot-scale vial freeze-drying process development

Samples are filled into vials and the stoppers are partially inserted before loading. The chamber freezes the samples and then creates a vacuum environment for sublimation drying. After the drying cycle is completed, the stoppering mechanism presses the stoppers into the vials inside the chamber. This reduces the need to expose dried products to ambient air before sealing.
|
Parameter |
LPJ-50Y |
LPJ-100Y |
LPJ-200Y |
|
Freeze-drying area (m²) |
0.5 |
1.15 |
2.2 |
|
Shelf size (mm) |
410 × 410 |
480 × 600 |
610 × 910 |
|
Shelf layers |
3+1 |
4+1 |
4+1 |
|
Shelf temperature range (°C) |
-55 to +70; optional ranges available |
||
|
Cold trap temperature (°C) |
≤ -75 / ≤ -85 / ≤ -110 (no load) |
||
|
Water trapping capacity (kg/24h) |
≥10 |
≥15 |
≥30 |
|
Installed power (kW) |
6.5 |
7.5 |
14.5 |
|
2 ml vial capacity (Ø16 mm) |
≥2,058 |
≥4,788 |
≥7,600 |
|
10 ml vial capacity (Ø22 mm) |
≥1,050 |
≥2,460 |
≥3,888 |
|
Overall dimensions (mm) |
1730 × 935 × 1855 |
1823 × 1144 × 2014 |
2193 × 1402 × 2370 |
|
Weight (kg) |
800 |
1000 |
1500 |
Available options include eutectic point testing, freeze-drying endpoint detection, vacuum pump anti-backflow protection, oil mist filtration, mobile APP monitoring, LYO-MEGA software, data recording, backfill filtration, automatic backfill and drainage control, capacitance vacuum measurement, and external vial hanging devices.
Please confirm the required options with LPMIE according to the intended process.
Choose the model according to the required shelf area, vial dimensions, batch size, and target vial quantity. Vial capacity varies with vial size and loading arrangement. Confirm vial compatibility before ordering.
1. What is the purpose of the stoppering system?
It presses vial stoppers into place inside the chamber after freeze-drying to help protect dried products from moisture and contamination.
2. Can the machine process bulk materials?
The equipment can support bulk and vial samples, depending on the selected configuration.
3. Can different vial sizes be used?
Common vial formats can be considered. Please provide vial dimensions and specifications for compatibility confirmation.
4. Does the machine support process monitoring options?
Yes. Options include endpoint detection, vacuum measurement, data recording, and monitoring software.
Contact LPMIE